Volume 3 Issue 3
Jul.  2010
Turn off MathJax
Article Contents
CHENG Shao-yuan, CAO Zhao-liang, HU Li-fa, MU Quan-quan, XUAN Li. Elimination of stray light reflected from anterior surface of cornea[J]. Chinese Optics, 2010, 3(3): 257-262.
Citation: CHENG Shao-yuan, CAO Zhao-liang, HU Li-fa, MU Quan-quan, XUAN Li. Elimination of stray light reflected from anterior surface of cornea[J]. Chinese Optics, 2010, 3(3): 257-262.

Elimination of stray light reflected from anterior surface of cornea

  • Received Date: 21 Feb 2010
  • Publish Date: 30 Jun 2010
  • The stray light reflected from the anterior surface of a cornea is about 10~100 times of the effective reflectivity of a retina, which can greatly reduce the precision of eye aberration detection and retinal imaging quality. Therefore, it is necessary to avoid the stray light. In this paper, some methods to eliminate the stray light reflected from the anterior surface of cornea are introduced and compared, such as the Polarization Beam Splitting(PBS), confocal hole, annular illumination, and the illumination with decentered thin beam. Analyzed results show that although the methods can remove the stray light reflected from the anterior surface of cornea in some degrees, they are not ideal. After analyzing, it is pointed out that the combination of annular illumination and confocal hole is the best, and the method not only has the high light-efficiency, but also is much more effective and handy than other methods.

     

  • loading
  • 加载中

Catalog

    通讯作者: 陈斌, bchen63@163.com
    • 1. 

      沈阳化工大学材料科学与工程学院 沈阳 110142

    1. 本站搜索
    2. 百度学术搜索
    3. 万方数据库搜索
    4. CNKI搜索
    Article views(4557) PDF downloads(1978) Cited by()
    Proportional views

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return