ZHANG Xiao-long, LIU Ying, SUN Qiang. Radiometric calibration of uncooled long-wave infrared thermal imager with high-precision[J]. Chinese Optics, 2012, 5(3): 235-241. doi: 10.3788/CO.20120503.0235
Citation: ZHANG Xiao-long, LIU Ying, SUN Qiang. Radiometric calibration of uncooled long-wave infrared thermal imager with high-precision[J]. Chinese Optics, 2012, 5(3): 235-241. doi: 10.3788/CO.20120503.0235

Radiometric calibration of uncooled long-wave infrared thermal imager with high-precision

  • Received Date: 21 Feb 2012
  • Rev Recd Date: 23 Apr 2012
  • Publish Date: 10 Jun 2012
  • In order to achieve the high accuracy temperature measurement for a uncooled long-wave infrared thermal imager under different ambient temperatures, the radiometric calibration model of the infrared thermal imager is established in consideration of the effect of detector operation temperature. The radiometric calibration experiment is then conducted, and sixteen functional relations between image gray levels and detector operation temperatures at different blackbody temperatures and different brightness gain values are obtained. The mathematical calculation model of target temperature is established and the calibration result is verified by experiments. The results show that the detector operation temperature effect can be treated linearly in the range from 25 ℃ to 40 ℃, and it is independent on the target temperature. By setting an appropriate brightness gain, the temperature measurement error of the infrared thermal imager is less than 0.5 ℃, which is a great improvement of the measurement accuracy of the uncooled long-wave infrared thermal imager under different ambient temperatures.

     

  • loading
  • [1] 李琼花,杨家德. 非致冷凝视红外焦平面阵列的特点及应用[J]. 重庆工学院学报 ,2003,17(1):48-50. LI Q H,YANG J D. The characteristics and applications of uncooled staring infrared focal plain array[J]. J. Chongqing Institute Technol.,2003,17(1):48-50.(in Chinese) [2] 孙恋君,张俊举,王世允,等. 非致冷微测辐射热计探测器工作温度特性研究[J]. 红外与毫米波学报 ,2007,26(3):232-236. SUN L J,ZHANG J J,WANG SH Y,et al.. Research on operating temperature characteristic of uncooled microbolometer detector[J]. J. Infrared Millim. Waves,2007,26(3):232-236.(in Chinese) [3] 黄英东,安建波. 基于改进多项式拟合的红外焦平面非均匀性校正方法[J]. 红外 ,2011,32(3):29-33. HUANG Y D,AN J B. A nonuniformity correction algorithm for IRFPA based on improved polynomial fitting[J]. Infrared,2011,32(3):29-33.(in Chinese) [4] DONG L Q,LIU X H,ZHAO Y J,et al.. Scene-based nonuniformity correction algorithm for MEMS-based un-cooled IR image system[J]. SPIE,2009,7419:74190Y. [5] ROBERT M P,DANIEL T P. Radiometric calibration of a longwave FPA camera for IR scene[J]. SPIE,2006,6208:680818. [6] WILLIAMS G M,BARTER A. Dual-band MWIR/LWIR radiometer for absolute temperature measurements[J]. SPIE,2006,6205:62050M. [7] 陆子凤. 红外热像仪的辐射定标和测温误差分析 .长春:中国科学院长春光学精密机械与物理研究所,2010. LU Z F. Calibration and measurement error analysis of infrared imaging system for temperature measurement . Changchun:Changchun Institute of Optics,Fine Mechanics and Physics,Chinese Academy of Sciences,2010.(in Chinese) [8] 乐逢宁,蔡静,马兰,等. 使用红外热像仪应注意的问题[J]. 计测技术 ,2010,30(增):100-101. YUE F N,CAI J,MA L,et al.. The problem should be attended when using IR thermal imager[J]. Calculate,Measure Technology,2010,30(s):100-101.(in Chinese) [9] 孙学金,刘剑,赵世军,等. 非致冷红外焦平面阵列的辐射定标模型[J]. 解放军理工大学学报 ,2008,9(4):399-403. SUN X J,LIU J,ZHAO SH J,et al.. Radiometric calibration model of uncooled infrared focal plane array[J]. J. PLA University of Science and Technology,2008,9(4):399-403.(in Chinese) [10] 张云涛,李涛,高太长. 基于人工神经网络的红外焦平面阵列定标方法[J]. 气象水文海洋仪器 ,2009(4):19-22. ZHANG Y T,LI T,GAO T CH. Calibration method of infrared focal plane arrays based on neural network[J]. Meteorological, Hydrological and Marine Instruments,2009(4):19-22.(in Chinese)